Capillary blood glucose (CBG) testing is one of the most common nursing procedures. It guides insulin doses, detects dangerous lows and highs, and helps manage diabetes, sepsis, liver disease, malnutrition, newborns and critically ill patients. Inaccurate results can lead to wrong insulin doses, so technique matters.
Units
Glucose is measured in mmol/L (UK, Europe, Australia, Canada, much of Africa) or mg/dL (US, India, the Gulf and others). To convert: mmol/L × 18 = mg/dL. For example, 4.0 mmol/L = 72 mg/dL; 10 mmol/L = 180 mg/dL.
Step-by-step technique
- Explain and gain consent; check identity.
- Check the meter: quality control done, strips in date and stored correctly, strip lot matches the meter if required.
- Ask the patient to wash their hands with warm water and dry thoroughly. Sugar on the fingers (from food or juice) gives falsely high readings. Alcohol wipes, if used, must dry fully.
- Hand hygiene and gloves.
- Use a single-use safety lancet on the side of the fingertip (less painful than the pad), rotating sites.
- Let a drop form; avoid excessive squeezing, which dilutes the sample with tissue fluid. Many guidelines advise wiping away the first drop.
- Apply the drop to the strip as instructed; read and record the result immediately, with the time.
- Dispose of the lancet in a sharps container; apply light pressure to the site.
Target ranges in hospital
Targets vary by guideline and patient. UK Joint British Diabetes Societies guidance commonly uses 6–10 mmol/L (108–180 mg/dL), with an acceptable range of 4–12 mmol/L, for most inpatients with diabetes. The American Diabetes Association suggests 140–180 mg/dL (7.8–10 mmol/L) for most hospitalised patients. Follow the target set for your patient.
- Below 4.0 mmol/L (72 mg/dL): treat hypoglycaemia. See our hypoglycaemia guide.
- Above 12–14 mmol/L (216–252 mg/dL): check ketones in people with type 1 diabetes or if unwell, and escalate. See our DKA guide.
When to check
- For eating patients on insulin, commonly before meals and at bedtime.
- For patients on IV insulin, hourly.
- Any patient with confusion, drowsiness, sweating, seizure or collapse.
- Patients on steroids, enteral or parenteral feeding, or with sepsis.
- Newborns at risk (small, preterm, babies of diabetic mothers), according to protocol.
Sources of error
- Sugar or other substances on the skin.
- Squeezing too hard.
- Expired or poorly stored strips (heat and humidity damage them, a real issue in hot climates).
- Poor peripheral circulation (shock, cold hands): capillary results may be inaccurate. Confirm with a venous sample.
- Severe anaemia or high haematocrit, and some medicines, can affect some meters.
- Taking blood from an arm with a glucose-containing drip running.
If a result doesn’t fit the patient’s condition, repeat it and, if in doubt, send a laboratory sample. Never delay treating a symptomatic hypoglycaemia while waiting.
Infection prevention
Shared meters and pen-type lancet devices have been linked to outbreaks of hepatitis B. Use single-use lancets only, and clean and disinfect shared meters between patients according to the manufacturer’s instructions.
Continuous glucose monitors
More patients now wear continuous glucose monitoring (CGM) sensors. Check your hospital’s policy on using CGM readings in hospital; finger-prick confirmation is often required for treatment decisions, especially for lows.
Based on Joint British Diabetes Societies inpatient guidance, American Diabetes Association Standards of Care and MHRA/CDC guidance on point-of-care glucose meters. Educational content: follow your local policy.